INMEL 60
This instrument has the possibility to generate and measure signals from ten thermocouples, five thermometer resistances, four voltage ranges, two current ranges, two resistance ranges. Thermoelectric terminals have an internal temperature measurement and for this reason make possible the generation or measurement of thermoelectric signals with the compensation for the given thermocouple. In this working mode, a real thermocouple is added to thermoelectric measuring terminals, and compensation conductors to thermoelectric generating terminals. The calibrator ensures the simultaneous generation of the signal on output terminals and the measurement of the signal on measuring terminals, thanks to this feature the calibrator is suitable to check transducers.
The INMEL 60 calibrator enables to check two-wire transducers.
The used graphical display contains a lot of information on a screen with a good readability. The display ground-lighting, which can be switched off, is especially useful in conditions of an insufficient lighting.
The internal supply from the battery enables the work not only in laboratory conditions but also on the object.
The calibrator can also be used to the maintenance and servicing of industrial measuring instruments, systems and electronic equipment.
The INMEL 60 calibrator enables to check two-wire transducers.
The used graphical display contains a lot of information on a screen with a good readability. The display ground-lighting, which can be switched off, is especially useful in conditions of an insufficient lighting.
The internal supply from the battery enables the work not only in laboratory conditions but also on the object.
The calibrator can also be used to the maintenance and servicing of industrial measuring instruments, systems and electronic equipment.
Checking and calibration:
The calibrator can also be used to the maintenance and servicing of industrial measuring instruments, systems and electronic equipment.
- measuring transducers controlled by a voltage or current signal,
- a thermoelement or resistance thermometer,
- two-wire transducers,
- temperature meters,
- pressure transducers,
- manometers,
- measuring and control lines.
The calibrator can also be used to the maintenance and servicing of industrial measuring instruments, systems and electronic equipment.
Basic parameters:
Range | Resolution | Accuracy1) | ||
Measurement | Generation | |||
V | VOLTAGE RANGES | |||
- 88mV up to 88 mV |
1 μV
|
± (0.05% M + 4d) | ± (0.05% S + 5μV) | |
- 220 mV up to 220 mV | 10 μV | ± (0.025% M + 3d) | ± (0.025% S + 30μV) | |
- 2.2 mV up to 2.2 mV | 100 μV | ± (0.025% M + 2d) | ± (0.025% S + 150μV) | |
- 11V up to 11 (measurement up to 25 V) | 1 mV | ± (0.025% M + 2d) | ± (0.025% S + 200μV) | |
I | CURRENT RANGES | |||
- 22 mA do 22 mA | 1 μV | ± (0.025% M + 3d) | ± (0.025% S + 3μA) | |
RESISTANCE RANGES | ||||
10 Ω up to 5000 Ω (measurement from 0Ω) | 0.1 Ω |
± (0.05% M + 2d)
Ip = 0.4 mA
|
± (0.05% S + 30 mΩ)
Ip = 10 mA
|
|
10 Ω up to 1000Ω (measurement from 0Ω) |
0.1 Ω |
± (0.05% M + 3d)
Ip = 0.4 mA
|
± (0.05% S + 150 mΩ)
Ip = 1 mA
|
|
RTD | RESISTANCE THERMOMETER RANGES | |||
-200°C up to 850°C (PT 100) | 0.1°C | ± (0.025% M + 0.3°C)Ip = 0.4 mA |
± (0.025% S +0,1°C)
Ip = 10 mA
|
|
-200°C up to 850°C (PT500, PT1000) |
± (0.025% S +0,1°C)
Ip = 1 mA
|
|||
-50°C up to 180°C (Cu 100) |
± (0.025% S +0,1°C)
Ip = 10 mA
|
|||
-60°C up to 180°C (Ni 100) |
± (0.025% S +0,1°C)
Ip = 10 mA
|
|||
T/C | THERMOCOUPLE RANGES | |||
J | -210°C up to 1200°C | 0.1°C | ± (0.025% M +0.5°C) | ± (0.025% S +0.5°C) |
K | -270°C up to 1370°C | ± (0.025% M +0.5°C) | ± (0.025% S +0.5°C) | |
T | -270°C up to 400°C | ± (0.025% M +0.5°C) | ± (0.025% S +0.5°C) | |
R |
-50°C up to 350°C
350°C up to 1769°C
|
± (0.05% M +1.5°C)
± (0.05% M +1°C)
|
± (0,05% S +1,5°C)
± (0,05% S +1°C) |
|
S |
-50°C up to 0°C
0°C up to 1769°C
|
0.1°C |
± (0.05% M +1.5°C)
± (0.05% M +1°C)
|
± (0,05% S +1,5°C)
± (0,05% S +1°C)
|
B | -200°C up to 1820°C | ± (0.025% M +1°C) | ± (0,025% S +1°C) | |
U | - 200°C up to 600°C | ± (0,025% M +0,5°C) | ± (0.025% S +0.5°C) | |
L | - 200°C up to 900°C | ± (0,025% M +0,5°C) | ± (0.025% S +0.5°C) | |
E | -50°C up to 700°C | ± (0,025% M +0,5°C) | ± (0.025% S +0.5°C) | |
N | - 270°C up to 1300°C | ± (0.025% M +0.5°C) | ± (0.025% S +0.5°C) | |
p | PRESSURE RANGES 2) | |||
- 100 up to 150 kPa
0 up to 7 MPa
|
±0.1%C M | ±0.1%C N |
Ip – measuring current
N- measurement value
M – setting value
1) During 12 months after the calibration, at temperature of 23± 5 °
2) Customer establishes the meauring range of the presuure transducer.
POWER SUPPLY | 230V +-10%, 50Hz, 14,3V/0,5A, 12VA |
BATTERY/POWER CONSUMPTION | 12V/ 1,9Ah/ 4VA max. |
DIMENSIONS | 230 x 90 x 205 mm |
WEIGHT | 2,2 kg |
OPERATION TIME | min. 4h |
BATTERY CHARGING TIME | 12 h |
INMEL 60 AS PRECISE THERMOMETER
- Measuring range : - 50 °C do 650°C
- Readout resolution : 0.01 °C
- Measuring sensor: PT 100
- with use standard sensor PT 100 : 0.02 % ± 0.1°C
- with use industrial sensor PT 100 with A class : 0.3 % ± 0.15°C
INMEL 60 CALIBRATOR with OPTION FOR PRESSURE CALIBRATION
INMEL 60 calibrator equipped in pressure transducer can serves to measurement of pressure value, and with pomp in set, can serves to calibration manometers and pressure transducers.
The range of pressure measurement determine the Customer, with place an order.
The error of measurement and pressure generation: +/- 0.1%
In set, in dependency from required pressure range, we offer:
1. pneumatic pomp TP1 - 40 – with pressure range from – 0.9 up to 40 bar
2. hydraulic pomp HTP1 - with pressure range from 0 up to 700 bar
The range of pressure measurement can be divided on two arbitrary, settled with user, sub ranges.
INMEL 60 calibrator equipped in pressure transducer can serves to measurement of pressure value, and with pomp in set, can serves to calibration manometers and pressure transducers.
The range of pressure measurement determine the Customer, with place an order.
The error of measurement and pressure generation: +/- 0.1%
In set, in dependency from required pressure range, we offer:
1. pneumatic pomp TP1 - 40 – with pressure range from – 0.9 up to 40 bar
2. hydraulic pomp HTP1 - with pressure range from 0 up to 700 bar
The range of pressure measurement can be divided on two arbitrary, settled with user, sub ranges.