Lider w produkcji kalibratorów
unia język polskijęzyk angielski

INMEL 52

Calibrator INMEL 52 integrate generation and measurement function of electric signals, used in automatic industry.
The main features of this devices is high accuracy and resolution of generating or measuring signals, friendly operation and large display.

Checking and calibration:
  • measuring transducers,
  • temperature meters,
  • measuring and control lines.

BASIC PARAMETERS

The measurement and generation

tabela1
Range
Resolution Measurement/simulation
Accuracy
4W ± Ω
Accuracy
± Ω
Measuring Current
mA
0.00Ω ~ 400.00Ω 0.01 / 0.1 Ω 0.1
0.15
0.1
0.1 ~ 0.5
0.5 ~ 3.0
400.0Ω ~ 1500.0Ω 0.1 / 0.1 Ω 0.5 0.5 0.05 ~ 0.8
1500.0Ω ~ 3200.0Ω 0.1 / 0.1 Ω
1
2
1 0.05 ~ 0.4

Admit excitation current only apply on simulate mode. The admit excitation current could be marked on the OHM meter or RTD meter which was connected to the calibrator. admit excitation current: 0.2mA. MAX input voltage: 30V. 2W/3W/4W – two wire/tree wire/four wire the resistance measuring set

Measure (input)/Simulate (output) RTD specification

The resolution of measurement and simulationi 0.1°C

 
 
 
tabela1
Mode Range Accuracy °C Measuring current mA
Measurement 4W  Measurement
2W/3W
Simulation
Pt10 385 -200~800°C / -328~1472°F 3.5 4.5 1.5 0.1~3.0
Pt50 385 -200~800°C / -328~1472°F 0.7 1.0 0.7 0.1~3.0
Pt100 385 -200~800°C / -328 ~ 1472°F 0.33 0.5 0.33 0.1~3.0
Pt200 385 -200~250°C / -328~482°F 0.2 0.3 0.2 0.1~3.0
250~630°C / 482~1166°F 0.8 1.6 0.8
Pt500 385 -200~500°C / -328~932°F 0.3 0.6 0.3 0.05~0.8
500~630°C / 932~1166°F 0.4 0.9 0.4
Pt1000 385 -200~100°C /-328~212°F 0.2 0.4 0.2 0.05~0.8
100~630°C /212~1166°F 0.2 0.5 0.2
Pt100 JIS -200~630°C /-328~1166°F 0.3 0.5 0.3 0.1~3.0
 

Admit excitation current only apply on simulate mode.
The admit excitation current could be marked on the OHM meter or RTD meter which was connected to the calibrator.
admit excitation current: 0.2mA.
MAX input voltage: 30V.
2W/3W/4W – two wire/tree wire/four wire the resistance measuring set


General Specifications:
  • Maximum voltage applied between any jack and earth ground or between any tow jack 30V
  • Resolution: RTD 0.1 °C/°F Resistance 0.01/0.1 Ω
  • Temperature coefficient ±0.01%/°C on 0°C 18°C and 28°C 50°C
  • Relative humidity: 95% to 30°C 75% to 40°C 45% to 50°C, 35% to 55°C
  • Safety: 1 meter drop test
  • Power requirements: 6 x AAA 1.5V Battery
  • Size: 205mm×98mm×46mm
  • Weight: 472 g (include battery)
     
Inquiry refer:
The offer should include: price
delivery time
payment terms
transport
training
warranty
certificate
the period of validity
Send The offer by: e-mail:
fax:
post:


*Name
*Department / Position
*Company Name
*Street
*Post code
*City
*Phone No
*E-mail
Comments
captcha
refresh text


zamknij
*Name
*Department / Position
*Company Name
*Street
*Post code
*City
*Phone No
*E-mail
Comments
captcha
refresh text


zamknij