INMEL 52
Calibrator INMEL 52 integrate generation and measurement function of electric signals, used in automatic industry.
The main features of this devices is high accuracy and resolution of generating or measuring signals, friendly operation and large display.
The main features of this devices is high accuracy and resolution of generating or measuring signals, friendly operation and large display.
Checking and calibration:
- measuring transducers,
- temperature meters,
- measuring and control lines.
BASIC PARAMETERS
The measurement and generation
Range |
Resolution Measurement/simulation
|
Accuracy
4W ± Ω
|
Accuracy
± Ω
|
Measuring Current
mA
|
0.00Ω ~ 400.00Ω | 0.01 / 0.1 Ω | 0.1 |
0.15
0.1
|
0.1 ~ 0.5
0.5 ~ 3.0
|
400.0Ω ~ 1500.0Ω | 0.1 / 0.1 Ω | 0.5 | 0.5 | 0.05 ~ 0.8 |
1500.0Ω ~ 3200.0Ω | 0.1 / 0.1 Ω |
1
2
|
1 | 0.05 ~ 0.4 |
Admit excitation current only apply on simulate mode. The admit excitation current could be marked on the OHM meter or RTD meter which was connected to the calibrator. admit excitation current: 0.2mA. MAX input voltage: 30V. 2W/3W/4W – two wire/tree wire/four wire the resistance measuring set
Measure (input)/Simulate (output) RTD specification
The resolution of measurement and simulationi 0.1°C
Mode | Range | Accuracy °C | Measuring current mA | ||
Measurement 4W | Measurement 2W/3W |
Simulation | |||
Pt10 385 | -200~800°C / -328~1472°F | 3.5 | 4.5 | 1.5 | 0.1~3.0 |
Pt50 385 | -200~800°C / -328~1472°F | 0.7 | 1.0 | 0.7 | 0.1~3.0 |
Pt100 385 | -200~800°C / -328 ~ 1472°F | 0.33 | 0.5 | 0.33 | 0.1~3.0 |
Pt200 385 | -200~250°C / -328~482°F | 0.2 | 0.3 | 0.2 | 0.1~3.0 |
250~630°C / 482~1166°F | 0.8 | 1.6 | 0.8 | ||
Pt500 385 | -200~500°C / -328~932°F | 0.3 | 0.6 | 0.3 | 0.05~0.8 |
500~630°C / 932~1166°F | 0.4 | 0.9 | 0.4 | ||
Pt1000 385 | -200~100°C /-328~212°F | 0.2 | 0.4 | 0.2 | 0.05~0.8 |
100~630°C /212~1166°F | 0.2 | 0.5 | 0.2 | ||
Pt100 JIS | -200~630°C /-328~1166°F | 0.3 | 0.5 | 0.3 | 0.1~3.0 |
Admit excitation current only apply on simulate mode.
The admit excitation current could be marked on the OHM meter or RTD meter which was connected to the calibrator.
admit excitation current: 0.2mA.
MAX input voltage: 30V.
2W/3W/4W – two wire/tree wire/four wire the resistance measuring set
- Maximum voltage applied between any jack and earth ground or between any tow jack 30V
- Resolution: RTD 0.1 °C/°F Resistance 0.01/0.1 Ω
- Temperature coefficient ±0.01%/°C on 0°C 18°C and 28°C 50°C
- Relative humidity: 95% to 30°C 75% to 40°C 45% to 50°C, 35% to 55°C
- Safety: 1 meter drop test
- Power requirements: 6 x AAA 1.5V Battery
- Size: 205mm×98mm×46mm
- Weight: 472 g (include battery)