The INMEL 21 calibrator is destined to check and control systems and industrial automatic control devices such as controllers, recorders, meters, etc. Moreover, it can be used for checking and adjustment of temperature measuring systems and also systems in which thermocouples and Pt 100 resistors are used..
The calibrator can be supplied from the mains or from a battery, which makes it suitable for mobile using.
The wide range temperature operation makes use of it in workshops.

The calibrator is equipped with :
  • 1 resistance range corresponding to Pt 100 resistor characteristic.
  • 1 direct voltage range of 10 V.
  • 2 direct current ranges of 5 mA and 20 mA.
  • 3 thermoelectric voltage ranges corresponding to thermocouples of the J [ Fe-CuNi], the K [ NiCr-NiAl], the S [PtRh10-Pt] type.

The three mode operation is possible at the J, K and S ranges :
  • the THCPL mode [ a thermocouple]: The voltage adequate to set temperature and a thermocouple type is given to an output of the calibrator, the instrument is connected through compensating leads to a measuring system in the place of a real thermocouple.
  • the SYSTEM 50°C mode [a thermocouple + a thermostat 50°C]: The voltage adequate to set temperature and a thermocouple type is given to the output of the calibrator assuming that a reference junction of a thermocouple exists at temperature of 50°C, the instrument is connected through two leads made of the same material to a measuring system in the place of a thermostat output signal.
  • the SYSTEM 0°C mode [a thermocouple + a thermostat 0°C] : The voltage adequate to set temperature and a thermocouple type is given to the output of the calibrator assuming that a reference junction of a thermocouple exists at temperature of 0°C, the instrument is connected through two leads made of the same material to an outside system.
APPLICATION:
  • measuring transducers controlled by a voltage or a current signal, a thermocouple or a resistance thermometer,
  • temperature meters,
  • measuring and control lines.

* we offer custom-made calibrator version with optionally chosen types of thermocouples

Basic parameters:

Voltage range parameters



tabela1
Parameter Value
Output voltage range
- 1V ... + 11V
- 0.5V...5.5V 4)
Resolution 0.01V
Accuracy 1) 0.08%S + 0.02%R
Additional error caused by change of ambient temperature 2) (0.02%S + 0.005%R) / 10°C
Additional error caused by change of load current 3) 0.01%R
Max. load current 22mA
Max. load reactance 10nF / 100mH
Short-circuit current 50 ... 80mA

1) 2) 3) S? set value, R ? range rated value 4) available in INMEL 20

Pt100 resistance range parameters


tabela1
Parameter Value
Temperature range - 200 ... + 850°C
Resolution 1°C
Accuracy 1) 0.08%S + 0.02%R
Additional error caused by change of ambient temperature2) (0.08%S + 0.02%R) / 10°C
Admissible measuring current (1 ... 5)mA

1) 2) S? set value, R ? range rated value

Current range parameters

tabela1
Parameter Range 5mA Range 20mA
Output current range -0.5mA ... +5.5mA -2mA ... +22mA
Resolution 0.001mA 0.01mA
Accuracy 1) 0.08%S + 0.02%R
Additional error caused by change of ambient temperature2) (0.02%S + 0.005%R) / 10°C
Additional error caused by change of load voltage 3) 0.01%R
Max. load voltage - 5.5V ... + 11V
Max. load reactance 10nF / 100mH
Off-load voltage <26V

1) 2) 3) S? set value, R ? range rated value


Thermoelectric range parameters


tabela1
Parameter Range J Range K Range S
Temperature range -210 ... +1200°C -270...+1372°C -50 ... +1769°C
Resolution 1°C
Accuracy1) 0.08%S + 0.02%R + 1°C
Additional error caused by change of ambient temperature2) (0.03%S + 0.01%R) / 10°C
Output resistance 3.5 Ω
Operation modes
THCPL (TERMOPARA)
SYSTEM 50°C
SYSTEM 0°C

1) 2) S? set value, R ? range rated value

General parameters:


tabela1
Max. power input from the mains 15VA
Max. power input from the battery 4W
Dimensions  (230 x 55 x 165)mm
Weight with the battery ok. 2.4kg
Work time of battery supply ok. 4h
Battery charging time ok. 12h



Inquiry refer:
The offer should include: price
delivery time
payment terms
transport
training
warranty
certificate
the period of validity
Send The offer by: e-mail:
fax:
post:


*Name
*Department / Position
*Company Name
*Street
*Post code
*City
*Phone No
*E-mail
Comments
captcha
refresh text


zamknij
*Name
*Department / Position
*Company Name
*Street
*Post code
*City
*Phone No
*E-mail
Comments
captcha
refresh text


zamknij